Sock Sizing Procedures for HIFOMACO LCS 4800
The procedures only apply to the HIFOMACO LCS 4800 model tested.
Calibration of the LCS 4800 should be maintained on an annual basis. Use the
supplied spring to verify that the machine is reading the same on a daily
basis.
Calibration Procedure
Calibration of the LCS 4800 is accomplished by the following procedure:
- Check the distance across the cross stretch arm fingers (Figure 1).
- Check the distance between the length stretch pins (Figure 2).
- If necessary loosen the set screws and adjust the pin distance to match
the cross stretch arm fingers distance (Figure 3).
- Adjust the reading on the display to match the distance read on both the
cross stretch arm fingers and length stretch pins (Figure 4).
- It may be necessary to zero the scale to match the arm setting (Figure
4a)
Daily Check of Machine
- Release the bottom carriage to travel the complete distance down
(Figure5)
- Raise the bottom carriage taking care not to allow the cable to catch
between the cross-stretch arms (Figure 6). If the cable becomes crimped
(Figure 6a), have it replaced.
- Each day confirm calibration of the machine by running 10 stretches with
the test spring supplied (Figure 7, 7a).
- If the readings are within the specified tolerance, begin your sock
measurements (Figure 7b).
- If adjustment is needed, call HIFOMACO @828-322-4292 for instructions.